CVE-2019-25719
HIGH SEVERITYCVSS Score & Metrics
Base Score
8.6 / 10
Vector String
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Vulnerability Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Vulnerability Details
Published Date
Last Modified
CWE ID
CWE-924
Source
NVD
Vendor
Dräger
Product
Infinity Acute Care System, Standalone Infinity M540 patient monitor