Back to CVE List

CVE-2019-25720

MEDIUM SEVERITY

CVSS Score & Metrics

Base Score
6.5 / 10
Vector String
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Vulnerability Description

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.

Vulnerability Details

Published Date
Last Modified
CWE ID
CWE-1286
Source
NVD
Vendor
Dräger
Product
SC 6002XL, SC6802XL, SC 7000, SC8000, SC90000 XL

External References

Discussion (0)

Add Comment

No comments yet. Be the first!