Back to CVE List

CVE-2021-4479

MEDIUM SEVERITY

CVSS Score & Metrics

Base Score
4.0 / 10
Vector String
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L

Vulnerability Description

Dräger Atlan A350 software versions 1.00 through 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.

Vulnerability Details

Published Date
Last Modified
CWE ID
CWE-1286
Source
NVD

External References

Discussion (0)

Add Comment

No comments yet. Be the first!