CVE-2025-15653

MEDIUM SEVERITY

CVSS Score & Metrics

Base Score
6.8 / 10
Vector String
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Vulnerability Description

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

Vulnerability Details

Published Date
Last Modified
CWE ID
CWE-668
Source
NVD
Vendor
Dräger
Product
Zeus IE, Zeus RS C500

External References