Back to CVE List

CVE-2026-7251

CRITICAL SEVERITY

CVSS Score & Metrics

Base Score
9.8 / 10
Vector String
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Vulnerability Description

Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Vulnerability Details

Published Date
Last Modified
CWE ID
CWE-259
Source
NVD

External References

Discussion (0)